• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAKO SURGICAL CORP. MCK TIBIAL ONLAY INSERT-SZ 5-8MM; PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAKO SURGICAL CORP. MCK TIBIAL ONLAY INSERT-SZ 5-8MM; PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM Back to Search Results
Catalog Number 180705-1
Device Problems Loss of Osseointegration (2408); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Discomfort (2330)
Event Date 09/19/2013
Event Type  Injury  
Manufacturer Narrative
An evaluation of the device cannot be performed as the device was not returned to the manufacturer due to hospital policy.Additional information has been requested.Should additional information become available at the later time it will be reported in a supplemental report upon completion of the investigation.
 
Event Description
Tibial loosening with mechanical failure, left uka.Revision of tibial compartment.
 
Manufacturer Narrative
An event regarding loosening involving an mako insert was reported.Conclusion: there is no indication that the product reported in this investigation contributed to the event since patient was revised due to tibial component loosening.The insert is an integral part of the tibial component construct and will have to be explanted together with the loose baseplate and no other allegations were made against the insert.No further investigation is required at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Event Description
Tibial loosening with mechanical failure, left uka.Revision of tibial compartment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MCK TIBIAL ONLAY INSERT-SZ 5-8MM
Type of Device
PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM
Manufacturer (Section D)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer (Section G)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer Contact
keyla colon
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key6439383
MDR Text Key71033805
Report Number3005985723-2017-00151
Device Sequence Number1
Product Code NPJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K090763
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type study
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/26/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date07/31/2015
Device Catalogue Number180705-1
Device Lot Number12200710
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/27/2017
Initial Date FDA Received03/28/2017
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received05/26/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/27/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other; Required Intervention;
Patient Age71 YR
Patient Weight87
-
-