Age at time of event: 18 years or older.Device is a combination product.(b)(4).Device evaluated by mfr.: a promus element mr ous 2.50 x 16 mm stent delivery system was returned for analysis.A visual examination of the stent found severe stent damage.Stent strut rows 1-6 from the proximal end of the stent appear undamaged; the remainder of the struts are damaged, stretched and pulled in a distal direction extending over the distal end of the tip.The stent was moved from the original crimped position in a distal direction 1.5 mm from the proximal markerband.The manufacturing crimp data for this device was reviewed.The maximum crimped stent profile was found to be within max crimped stent profile measurement.The bumper tip of the device was examined and no issues were noted.The balloon body was reviewed and no issues were noted with the overall balloon.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and tactile examination of the hypotube profile found no issues.An examination of the shaft polymer extrusion profile found no issues.The bi-component bond showed no signs of damage or strain.No other issues were identified during the product analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The root cause is not confirmed as there was no evidence of the alleged issue or any anomalies which could have contributed to the reported difficulty.(b)(4).
|
Reportable based on analysis completed on 14mar2017.It was reported that shaft kink occurred.The target lesion was located in a coronary artery.A 2.50 x 16 mm promus element ¿ stent was advanced to treat the lesion.However, during procedure, it was noted that the delivery shaft was kinked.No patient complications were reported and the patient's status was stable.However, returned device analysis revealed stent moved on balloon and stent damaged.
|