BIOMET ORTHOPEDICS COMPREHENSIVE REVERSE SHOULDER HUMERAL TRAY WITH LOCKING RING 44MM STANDARD; PROSTHESIS, SHOULDER Back to Search Results
Catalog Number 115340
Device Problem Fracture (1260)
Patient Problem No Information (3190)
Event Date 02/28/2017
Event Type  Injury  
Manufacturer Narrative
(b)(4).Customer has indicated that the product will not be returned to zimmer biomet for investigation as the hospital has discarded the implant.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It is reported the patient underwent shoulder arthroplasty.Subsequently, post-operative x-ray on unknown date revealed the humeral tray had fractured.Patient underwent revision procedure approximately twenty-two months post-implantation due to the fractured taper on the humeral tray.No additional patient consequences were reported.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Medical product: comprehensive nano humeral stem, catalog#: 115740, lot#: 951520.Reported event was able to be confirmed from a photograph received, showing the device taper to be fractured.Device history record (dhr) was reviewed and no discrepancies relevant to the reported event were found.Review of the complaint history determined that no further action(s) is/are required.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMPREHENSIVE REVERSE SHOULDER HUMERAL TRAY WITH LOCKING RING 44MM STANDARD
Type of Device
PROSTHESIS, SHOULDER
Manufacturer (Section D)
BIOMET ORTHOPEDICS
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
BIOMET ORTHOPEDICS
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key6440039
MDR Text Key71059260
Report Number0001825034-2017-02174
Device Sequence Number1
Product Code PAO
Combination Product (y/n)N
Reporter Country CodeKS
PMA/PMN Number
PK080642
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/18/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/28/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Physician
Device Catalogue Number115340
Device Lot Number945480
Other Device ID NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/15/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/17/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-