Results: the ruby coil introducer sheath was kinked approximately 12.0, 42.0, 43.0, and 54.0 cm from its proximal end.There were mild bends in the pusher assembly approximately 10.0 and 22.0 cm from its proximal end.Conclusions: evaluation of the device revealed a kinked introducer sheath.This type of damage may result from forcefully advancing the device at extreme angles into the rhv.Further evaluation revealed slight bends in the pusher assembly.This damage is likely incidental and may have occurred during packaging for return to penumbra facilities.Penumbra coils are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
The patient was undergoing a coil embolization procedure in the splenic artery using ruby coils.During the procedure, the hospital staff noticed that the a ruby coil introducer sheath was bent upon insertion into the rhv on the non-penumbra microcatheter.Consequently the ruby coil was unable to advance within the introducer sheath and therefore, it was removed, and the procedure was completed using additional ruby coils.There was no report of an adverse effect to the patient.
|