The customer reported that the transmitter displayed incorrect ecg readings while on a simulator.During testing, the device displayed only noise/artifact.They tried new lead sets; however the issue continued to persist.The simulator's heart rate was set to read 80 beats per minute, but it displayed a reading of 90 beats per minute.The customer will send the device in for evaluation.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
|
(b)(4).Manufacturer narrative: the customer reported that the transmitter displayed incorrect ecg readings while on a simulator.During testing, the device displayed only noise/artifact.They tried new lead sets; however, the issue continued to persist.The simulator's heart rate was set to read 80 beats per minute, but it displayed a reading of 90 beats per minute.The reported problem could not be duplicated.In addition, no fluid intrusion was observed on the main board.However, there is significant physical damage that will require replacement of the case.The customer was sent an exchange unit.Nihon kohden will submit a supplemental report in accordance with 21 cfr part 803.56 if additional information becomes available.
|