• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NIHON KOHDEN CORPORATION ZM-531PA; TRANSMITTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NIHON KOHDEN CORPORATION ZM-531PA; TRANSMITTER Back to Search Results
Model Number ZM-531PA
Device Problem Device Alarm System (1012)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/10/2017
Event Type  malfunction  
Manufacturer Narrative
The customer reported that the transmitter did not alarm asystole when it should have.The alarm was set to three seconds but when tested on two simulators, they both took ten seconds for the asystole alarm to trigger.Alarm and arrhythmia default settings were reviewed by a nihon kohden clinician.The asystole was set for 3 seconds and the pause for 2 seconds.Alarm and recall were both turned on for warning level alarms.When asked about viewing the event in full disclosure, the nurse stated that the patient had been transferred and there was no available data.No logs or patient data were available for nihon kohden to assess.A follow up call to the hospital was made to see if similar issues have occurred.Customer states that they have not experienced any similar occurrences.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.Device not returned.
 
Event Description
The customer reported that the transmitter did not alarm asystole when it should have.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZM-531PA
Type of Device
TRANSMITTER
Manufacturer (Section D)
NIHON KOHDEN CORPORATION
1-31-4 nishiochia, shinjuku-ku
attn: shama mooman
tokyo, japan 161-8 560
JA  161-8560
Manufacturer (Section G)
NIHON KOHDEN TOMIOKA CORPORATION
1-1 tajino
attn: shama mooman
tomioka city, gunma 370-2 314
JA   370-2314
Manufacturer Contact
shama mooman
safety mgmt dept, quality mgmt
seibu bldg 2, 4th floor 1-11-2
kusunokidai tokorozawa, saitama 359-8-580
JA   359-8580
2687488
MDR Report Key6462302
MDR Text Key71836540
Report Number8030229-2017-00099
Device Sequence Number1
Product Code DRT
UDI-Device Identifier04931921115107
UDI-Public04931921115107
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K043517
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 04/05/2017,03/10/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberZM-531PA
Device Catalogue NumberZM-531PA
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA04/05/2017
Distributor Facility Aware Date03/10/2017
Device Age35 MO
Event Location Hospital
Date Report to Manufacturer04/05/2017
Initial Date Manufacturer Received 04/05/2017
Initial Date FDA Received04/05/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/08/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-