Model Number DTMA2QQ |
Device Problem
Pacing Problem (1439)
|
Patient Problem
Palpitations (2467)
|
Event Date 01/26/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Product event summary - the device was not returned for analysis.However, performance data collected from the device was received and analyzed.Analysis of the device memory indicated that the ventricular safety pacing feature was causing non-sustained runs of ventricular tachycardia.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the patient experienced palpitations.The ventricular safety pacing in ventricular sensed was causing ventricular tachycardia episodes.The device was reprogrammed (ventricular safety pacing was turned off) and remains in use.No further patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
This device was reported as included in the field action noted but returned product investigation found the device did not perform as described in the field action.The device is no longer included as part of the field action.If information is provided in the future, a supplemental report will be issued.
|
|