Model Number DTBA1D1
Device Problems High impedance (1291); Misconnection (1399); Protective Measures Problem (3015)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/30/2016
Event Type  Injury  
Manufacturer Narrative
Concomitant medical products: 5076-45 lead, implanted: (b)(6) 2005.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the patient presented to the clinic approximately two months following implant of the device due to an alert from the cardiac resynchronization therapy defibrillator (crt-d) for high right ventricular (rv) lead impedance on a competitor lead.The patient was scheduled for a lead revision due to a possible crt-d set screw issue.The crt-d remains in use.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIVA XT
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key6470532
MDR Text Key72066395
Report Number3004209178-2017-07781
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00643169530140
UDI-Public00643169530140
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 01/04/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/06/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date03/14/2018
Device Model NumberDTBA1D1
Device Catalogue NumberDTBA1D1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/04/2017
Date Device Manufactured09/22/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
7122 LEAD
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age86 YR
-
-