Siemens was notified on november 30, 2016 that the image quality on the oncor impression plus system was bad and siemens began servicing the system to resolve the issue.On march 13, 2017, the customer provided siemens with new information in regard to dose measurements of the affected system that clearly points to a mis-calibrated dose monitor system when using the film mode.The control console settings on the "low dose page" for 6 mv photons were checked and found to be incorrect.As a result, images for patient position verification using the film mode have been acquired with a higher dose than intended by the user.On march 14, 2017 the user was advised to discontinue the clinical use of the film mode and treatment techniques using the low dose page.The customer has reported that 50 patients have received a higher dose than intended during image acquisition however there is no report of injury.This reported issue occurred in (b)(6).
|
The service engineer with the site physicist checked all modes.The low dose page and the film mode have been re-calibrated as of march 16, 2017.The user manual has been updated with detailed information with regards to the calibration of the film mode and has been distributed to the installed base.Considering this, no further corrective action is initiated.
|