Model Number 1458Q/86 |
Device Problems
Fracture (1260); High impedance (1291); No Pacing (3268)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 03/14/2017 |
Event Type
Injury
|
Manufacturer Narrative
|
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
|
Event Description
|
Related manufacturer report number: 2017865-2017-02228.There was high impedance with no pacing on the right atrial (ra) and left ventricular (lv) leads due to conductor fractures.Both leads were explanted and replaced and the patient was stable.
|
|
Manufacturer Narrative
|
As received, a partial lead (connector end) was returned for evaluation in one piece.Visual inspection of the lead revealed a breach of all the lumens, fracturing all the cables and all four filars of the inner coil distal to the suture sleeve tie impression.The cause of the reported events was consistent with clavicle crush.
|
|