Model Number P1170E0000051
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/31/2017
Event Type  malfunction  
Manufacturer Narrative
The hill-rom technician found the communication cable needed to be replaced.Per the hill-rom service manual the bed should be subject to an effective maintenance program.An annual service of the bed is advised in order to maintain its characteristics and performance.Sidecom® communication system: inspect and test the communication junction box.Make sure the sidecom® communication system features operate correctly.Inspect the communication cable, including the male and female pins in the plug.Replace as necessary.A search of the hill-rom maintenance records showed hill-rom performed preventative maintenance on this bed in 2016.It is unknown if the facility performed any other preventative maintenance on this bed.The technician replaced the communication cable to resolve the issue.Based on this information, no further action is required.
 
Event Description
Hill-rom received a report from the account stating the bed exit was alarming, but not sending a priority call.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint #(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAREASSIST ES130/230/430, NUL
Type of Device
A/C POWERED ADJUSTABLE HOSPITAL BED
Manufacturer (Section D)
HILL-ROM MEXICO
ave.del telefono no. 200 col. huinala
apodaca, nuevo leon 66640
MX  66640
Manufacturer Contact
erin padgett
1069 state route 46 east
batesville, IN 47006
8129312791
MDR Report Key6494027
MDR Text Key73025733
Report Number3006697241-2017-00048
Device Sequence Number1
Product Code FNL
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial
Report Date 03/31/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/14/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other Caregivers
Device Model NumberP1170E0000051
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received03/31/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/16/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-