The insulin pump alarmed motor error during rewind due to motor encoder signal out of phase.Unable to perform self-test, off no power test, unexpected restart error test, occlusion test, prime test, excessive no delivery test, displacement test due to motor error alarm.The insulin pump passed idle current test and run current test.Unable to verify motor position encoder error alarm and check setting alarm due to motor error alarm.The drive support was inspected and no anomaly was noted.The insulin pump had cracked reservoir tube lip.
|