• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP; INSULIN INFUSION PUMP / SENSOR AUGMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP; INSULIN INFUSION PUMP / SENSOR AUGMENTED Back to Search Results
Model Number MMT-723NAP
Device Problems Device Displays Incorrect Message (2591); Positioning Problem (3009)
Patient Problem Hyperglycemia (1905)
Event Date 04/06/2017
Event Type  malfunction  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
The customer called to report that she wore the insulin pump during an mri scan and afterwards the device alarmed motor error and motor position encoder error.Customer's blood glucose reading was 195 mg/dl.Customer was able to rewind the device.The device settings indicated that there was a history of other motor error alarms.The customer was advised to discontinue the device and revert to a backup plan.Device was replaced and will be returned for analysis.
 
Manufacturer Narrative
The insulin pump alarmed motor error during rewind due to motor encoder signal out of phase.Unable to perform self-test, off no power test, unexpected restart error test, occlusion test, prime test, excessive no delivery test, displacement test due to motor error alarm.The insulin pump passed idle current test and run current test.Unable to verify motor position encoder error alarm and check setting alarm due to motor error alarm.The drive support was inspected and no anomaly was noted.The insulin pump had cracked reservoir tube lip.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Type of Device
INSULIN INFUSION PUMP / SENSOR AUGMENTED
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer Contact
gerwin de graaff
18000 devonshire street
northridge, CA 91325-1219
8185764805
MDR Report Key6496258
MDR Text Key73102846
Report Number3004209178-2017-43990
Device Sequence Number1
Product Code OYC
Combination Product (y/n)N
PMA/PMN Number
P980022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial,Followup
Report Date 05/12/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-723NAP
Device Catalogue NumberMMT-723NAP
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/14/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/06/2017
Initial Date FDA Received04/15/2017
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received05/26/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age62 YR
Patient Weight100
-
-