ABBOTT DIAGNOSTICS DIVISION CELL-DYN RUBY ANALYZER; AUTOMATED HEMATOLOGY ANALYZER Back to Search Results
Catalog Number 08H67-01
Device Problem High Test Results (2457)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/27/2017
Event Type  malfunction  
Manufacturer Narrative
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.
 
Event Description
The customer stated that the cell dyn ruby did not generate rrbc flagging on one sample when run in the cbc mode.The initial wbc result on (b)(6) 2017 run in cbc mode - sid 15ad= 58,800 ul/ repeated in cbc+rrbc mode wbc = 16,300 ul.The patient is a known sickle cell.There was no addiitonal patient information provided.
 
Manufacturer Narrative
The investigation consisted of review of submitted data, review of product historical data, and product labeling.A review of product historical data did not identify any trends or increase in complaint activity.A review of the customer's submitted data showed the resistant rbcs may have been misclassified as lymphocytes in the initial run in cbc test selection.Repeat testing in cbc + rrbc test selection generated the accepted wbc result with no flagging.A review of the cell-dyn ruby system operator's manual provides descriptor/flag - cause and suggested action.The manual also states that nwbc flag suggests the presence of lytic-resistant rbc and if accompanied with another flag that manual smear review is suggested.A malfunction was identified as the device failed to meet performance specifications or otherwise perform as intended at the customer site.However, a systemic issue and/or product deficiency was not identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELL-DYN RUBY ANALYZER
Type of Device
AUTOMATED HEMATOLOGY ANALYZER
Manufacturer (Section D)
ABBOTT DIAGNOSTICS DIVISION
4551 great america parkway
santa clara CA 95054
Manufacturer (Section G)
ABBOTT DIAGNOSTICS DIVISION
4551 great america parkway
santa clara CA 95054
Manufacturer Contact
noemi romero-kondos, rn bsn
100 abbott park road
dept. 09b9, lccp1-3
abbott park, IL 60064-3537
224667-512
MDR Report Key6500350
MDR Text Key73073901
Report Number2919069-2017-00071
Device Sequence Number1
Product Code GKZ
UDI-Device Identifier00380740017170
UDI-Public00380740017170
Combination Product (y/n)N
Reporter Country CodeBR
PMA/PMN Number
K061667
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 05/02/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number08H67-01
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/27/2017
Initial Date FDA Received04/18/2017
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received05/02/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/30/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Age25 YR
-
-