• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC. UNIFY CRT-D; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC. UNIFY CRT-D; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD3235-40
Device Problem Programming Issue (3014)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/30/2017
Event Type  malfunction  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
During follow up, inappropriate automatic mode switch episodes and competitive atrial pacing occurred due to previous reprogramming changes.New reprogramming recommendations were given to the physician.The patient is in stable condition and did not receive any inappropriate therapy due to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIFY CRT-D
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
645 almanor avenue
sunnyvale CA 94085
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
645 almanor avenue
sunnyvale CA 94085
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key6501196
MDR Text Key73076342
Report Number2938836-2017-22103
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 04/18/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2012
Device Model NumberCD3235-40
Device Catalogue NumberCD3235-40
Device Lot Number3254775
Other Device ID Number05414734503556
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/30/2017
Initial Date FDA Received04/18/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/03/2010
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-