MAUDE Adverse Event Report: AV-TEMECULA-CT XIENCE ALPINE EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT XIENCE ALPINE EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT SYSTEM Back to Search Results
Catalog Number 1125325-23
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Atrial Fibrillation (1729); Non specific EKG/ECG Changes (1817)
Event Date 11/30/2016
Event Type  Injury  
Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device was not returned for evaluation.Based on the case information and related record review, a conclusive cause for the reported patient effect(s), and the relationship to the product, if any, cannot be determined.However, atrial arrhythmia is listed in the xience alpine everolimus eluting coronary stent systems instructions for use (ifu) procedures.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Event Description
It was reported that the index procedure took place on (b)(6) 2016 during which a 3.25 x 23 mm xience alpine was placed for treatment.The patient was readmitted to the hospital on (b)(6) 2016 with abnormal ekg/ecg and experiencing symptoms of atrial fibrillation.Although the treatment performed is unspecified, the fact that rehospitalization occurred indicates that some form of treatment was performed.No other information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE ALPINE EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Type of Device
DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
CLONMEL, IRELAND REG# 9616693
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key6501443
MDR Text Key73100057
Report Number2024168-2017-03298
Device Sequence Number1
Product Code NIQ
UDI-Device Identifier08717648199455
UDI-Public(01)08717648199455(17)190928(10)6092141
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P110019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date 04/18/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date09/28/2019
Device Catalogue Number1125325-23
Device Lot Number6092141
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/27/2017
Initial Date FDA Received04/18/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/01/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age61 YR
Patient Weight99
-
-