(b)(6) 2017 right depth lead replacement right mesial temporal.(b)(6) 2016 patient had scheduled neurostimulator replacement.At the time of replacement, rh (lead 1 sn (b)(4)) impedances were >3000 ohms or "insufficient charge".Troubleshooting was performed including; cleaning leads, removing connector cover and irrigating, changing connector covers, swapping leads - there was no change to lead impedance measurements from that lead.Also tried plugging in the old device, but same results as with the new device.Real time ecog looked fine, so closed.(b)(6) 2016 follow up visit.Detailed therapy testing on the right hippocampal lead was completed and patient reported no sensitivity.Palpitation near the lead was done while running real-time ecog, and there was some artifact present in the ecog.The device was reprogrammed to remove the stimulation from lead 1 (rh).(b)(6) 2017 the right depth lead was replaced (sn (b)(4), right mesial temporal) to address the high impedances.The lead was discarded by the site.
|