MAUDE Adverse Event Report: GUIDANT CRM CLONMEL IRELAND INOGEN; IMPLANTABLE CHF GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GUIDANT CRM CLONMEL IRELAND INOGEN; IMPLANTABLE CHF GENERATOR Back to Search Results
Model Number G148
Device Problems Over-Sensing (1438); Pacing Problem (1439); Ambient Noise Problem (2877)
Patient Problem Complete Heart Block (2627)
Event Date 03/22/2017
Event Type  Injury  
Manufacturer Narrative
As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
 
Event Description
Boston scientific received information that following implant of this system, a review of telemetry strips noted three seconds pauses in pacing which resulted in asystole for this patient.A boston scientific technical services consultant reviewed the electrograms and discussed the resemblance to crosstalk from the atrial paced events on the right ventricular (rv) channel.The consultant confirmed the signal is not being oversensed in most complexes.The boston scientific sales representative stated an x-ray and all lead measurements were stable and consistent with implant numbers.The physician kept the patient an extra night in the hospital for observation.System evaluation the following day showed the clinical observations were still occurring intermittently and the physician suspects it is due to the proximity of the rv coil to a valve.Reprogramming was performed and the patient was going to be seen again for defibrillation threshold testing (dft).No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INOGEN
Type of Device
IMPLANTABLE CHF GENERATOR
Manufacturer (Section D)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer (Section G)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer Contact
tim degroot
4100 hamline ave. n
st. paul, MN 
6515824786
MDR Report Key6504144
MDR Text Key73218707
Report Number2124215-2017-06241
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526534591
UDI-Public(01)00802526534591(17)20180927
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 03/23/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date09/27/2018
Device Model NumberG148
Other Device ID NumberINOGEN X4 CRT-D
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Initial Date Manufacturer Received 03/23/2017
Initial Date FDA Received04/19/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/27/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
0275; 4671; 7740; G148
Patient Outcome(s) Hospitalization; Life Threatening; Other;
Patient Age77 YR
-
-