MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE Back to Search Results
Model Number 8637-20
Device Problem Failure to Interrogate (1332)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/10/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a foreign manufacturer representative regarding a pump that was never implanted.There was no indication of patient interaction.There was no drug in the pump.Pump interrogation was attempted prior to surgery, but interrogation was not successful.There were no known environmental/external/patient factors that may have led or contributed to the issue.No diagnostics/troubleshooting was able to be performed.A back up pump was used instead.The issue was considered resolved as of (b)(6) 2017.The issue was considered to be an out-of-box failure.The pump would be returned to the manufacturer.No complications were reported/anticipated.
 
Manufacturer Narrative
Analysis of the pump (sn; (b)(4) found no anomaly.The pump was returned and passed all functional testing in the laboratory.Per the complaint of difficult telemetry standard telemetry distance test was performed with typical results, pass.(b)(4).The previously reported (b)(4) no longer applies to this event.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCHROMED II
Type of Device
PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key6504647
MDR Text Key73252140
Report Number3004209178-2017-08624
Device Sequence Number1
Product Code LKK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P860004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/19/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/19/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/14/2018
Device Model Number8637-20
Device Catalogue Number8637-20
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/24/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received06/09/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/23/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-