• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC, INC. OPEN PIVOT AP360 HEART VALVE; HEART-VALVE, MECHANICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC, INC. OPEN PIVOT AP360 HEART VALVE; HEART-VALVE, MECHANICAL Back to Search Results
Model Number 505D
Device Problem Mechanical Jam (2983)
Patient Problems Corneal Pannus (1447); Thrombus (2101)
Event Date 03/03/2017
Event Type  Injury  
Manufacturer Narrative
Product analysis: the device has not been returned for evaluation.Conclusion: based on the received information the leaflet motion was caused by thrombus formation.This investigation revealed no intrinsic evidence to suggest a root cause related to manufacturing.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information that this mechanical valve was implanted in the mitral position with a 20 mm gore-tex graft skirting the valve to prevent against pannus formation.Following implant, a leaflet dysfunction was observed, presumably resulting from pannus on the graft.Approximately six months post-implant of this device, surgical intervention was undertaken to remove a thrombus attached to the leaflet and correct the observed leaflet dysfunction; the device was not replaced.No other adverse patient effects were reported.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPEN PIVOT AP360 HEART VALVE
Type of Device
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
MEDTRONIC, INC.
3800 annapolis lane
minneapolis MN 55447
Manufacturer (Section G)
MEDTRONIC, INC.
3800 annapolis lane
minneapolis MN 55447
Manufacturer Contact
paula bixby
8200 coral sea street ne
mounds view, MN 55112
7635055378
MDR Report Key6505182
MDR Text Key73215302
Report Number3008592544-2017-00020
Device Sequence Number1
Product Code LWQ
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P990046
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/05/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/19/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date11/05/2020
Device Model Number505D
Device Catalogue Number505DM16
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received04/05/2017
Was Device Evaluated by Manufacturer? No
Date Device Manufactured11/07/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age1 YR
Patient Weight4
-
-