• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: W.L. GORE & ASSOCIATES GORE® CARDIOFORM SEPTAL OCCLUDER; OCCLUDER, TRANSCATHETER SEPTAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W.L. GORE & ASSOCIATES GORE® CARDIOFORM SEPTAL OCCLUDER; OCCLUDER, TRANSCATHETER SEPTAL Back to Search Results
Catalog Number GSX0030A
Device Problem Migration or Expulsion of Device (1395)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/24/2017
Event Type  Injury  
Manufacturer Narrative
A review of the manufacturing records for the device verified that the lot met all pre-release specifications.The device was discarded, so no engineering investigation could be performed.(b)(4).
 
Event Description
It was reported the physician was implanting a gore® cardioform septal occluder to close an atrial septal defect.The approximate size of the defect was 16mm, no balloon measurement was taken.The device was deployed with good apposition to the septum.A couple of hours post-procedure the device embolized to the aortic arch.The physician retrieved the device with a snare.The patient was doing well following the procedure and will be brought back at a later date for defect closure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® CARDIOFORM SEPTAL OCCLUDER
Type of Device
OCCLUDER, TRANSCATHETER SEPTAL
Manufacturer (Section D)
W.L. GORE & ASSOCIATES
flagstaff AZ
Manufacturer (Section G)
KENDRICK PEAK MPD B/P
4250 w. kiltie lane
flagstaff AZ 86005
Manufacturer Contact
ashley marostica
1500 n. 4th street
flagstaff, AZ 
9285263030
MDR Report Key6505285
MDR Text Key73216177
Report Number2017233-2017-00210
Device Sequence Number1
Product Code MLV
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P050006
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/24/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date11/17/2018
Device Catalogue NumberGSX0030A
Device Lot Number15704840
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received04/19/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/17/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age39 YR
Patient Weight120
-
-