• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP; INSULIN INFUSION PUMP / SENSOR AUGMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP; INSULIN INFUSION PUMP / SENSOR AUGMENTED Back to Search Results
Model Number MMT-723NAB
Device Problems Break (1069); Positioning Problem (3009)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/18/2017
Event Type  malfunction  
Manufacturer Narrative
Pump received with minor scratched lcd window, broken reservoir tube lip, broken belt clip slot and cracked battery tube threads.A test reservoir was installed and did not lock in place due to complete broken reservoir tube lip.
 
Event Description
The customer reported via phone call that the plastic transparent ring at the reservoir compartment is broken and the reservoir cannot stay in place.Customer's blood glucose was 125 mg/dl at the time of incident.Customer was advised to discontinue use of the device and revert to a back-up plan per their doctor's instruction.The customer was also advised that the device would be replaced and agreed to return the product for analysis.No further details given.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Type of Device
INSULIN INFUSION PUMP / SENSOR AUGMENTED
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer Contact
gerwin de graaff
18000 devonshire street
northridge, CA 91325-1219
8185764805
MDR Report Key6505895
MDR Text Key73336235
Report Number3004209178-2017-44539
Device Sequence Number1
Product Code OYC
Combination Product (y/n)N
PMA/PMN Number
P980022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial
Report Date 03/28/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-723NAB
Device Catalogue NumberMMT-723NAB
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/01/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/28/2017
Initial Date FDA Received04/19/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age27 YR
Patient Weight54
-
-