TERUMO CARDIOVASCULAR SYSTEMS CORP. ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
|
Back to Search Results |
|
Model Number 149673 |
Device Problems
Failure To Adhere Or Bond (1031); Break (1069)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 03/29/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
It was reported that during set-up of the device for a cardiopulmonary bypass procedure, the latch in the air bubble detector was broken.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
|
|
Manufacturer Narrative
|
The user facility biomedical technician (bmet) mentioned that they are going to order new latches for their air bubble detector (abd) to resolve the issue.
|
|
Manufacturer Narrative
|
The reported complaint was confirmed via the user facility's biomedical technician (bmet) observation, as broken latch is a readily identifiable condition.Per the bmet, the issue was resolved by replacing latch/lock.No parts were returned to the manufacturer for evaluation.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|