MICROMAR INDUSTRIA E COMERCIO LTDA EASYDRILL PERFORATOR; DRILLS, BURRS, TREPHINES & ACCESSORIES (COMPOUND, POWERED)
|
Back to Search Results |
|
Model Number DM0010FAA |
Device Problem
Device Remains Activated (1525)
|
Patient Problems
Contusion (1787); Perforation (2001)
|
Event Date 03/22/2017 |
Event Type
Injury
|
Manufacturer Narrative
|
Report inconclusive.No evaluation was performed, as the device was not returned.If the device is returned in the future, product analysis may be performed.This dm0010faa perforator with lot number 720/16 was manufactured by micromar.Multiple warnings are included in the easydrill cranial perforator ifu manual including: it is essential to keep the drill perpendicular (90°) at the predetermined point of the skull to be drilled, as an excessive deviation from perpendicularity may cause the product to fail and lead to serious patient injury.Select a drill bit suitable for the bone thickness.This prevents the drill from tearing the bone or brain tissue (similar effect when the bit is not positioned at 90°).If the components of the easydrill cranial perforator become loose during trephination, discontinue use.The drill can cut or tear the dura mater when it unlocks.Check some conditions before performing the procedure, such as the existence of adhering dura mater or other anomalies adjacent to the drilling site.We will continue to monitor this complaint type for trends.Additional importer medwatch information: (b)(4).A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the perforator tool did not stop rotating which caused an injury reported as "lesions of the dura and cerebellum contusion." it was indicated the patient outcome was ok, and alive with the reported injury.The surgeon repaired the dura defect and hemostasis of cerebellum.On follow up it was confirmed the neurosurgeon was performing trephination with the perforator when the reported event details occurred.However, it was confirmed the patient does not require follow up due to the event.This was the initial use of the perforator as well as the first perforation.Partial concomitant information was provided.
|
|
Manufacturer Narrative
|
Report inconclusive.This dm0010faa perforator with lot number 720/16 was manufactured by micromar.The perforator has been returned to the manufacturer and is still pending their evaluation results.The device history records were reviewed by micromar and all specifications were met prior to release.Multiple warnings are included in the easydrill cranial perforator ifu manual including: it is essential to keep the drill perpendicular (90°) at the predetermined point of the skull to be drilled, as an excessive deviation from perpendicularity may cause the product to fail and lead to serious patient injury.Select a drill bit suitable for the bone thickness.This prevents the drill from tearing the bone or brain tissue (similar effect when the bit is not positioned at 90°).If the components of the easydrill cranial perforator become loose during trephination, discontinue use.The drill can cut or tear the dura mater when it unlocks.Check some conditions before performing the procedure, such as the existence of adhering dura mater or other anomalies adjacent to the drilling site.We will continue to monitor this complaint type for trends.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
No conclusion can be drawn.Evaluation for continuing to spin could not be performed because the perforator was damaged.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|