Model Number 466FXXXX |
Device Problems
Occlusion Within Device (1423); Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Occlusion (1984); Thrombosis (2100)
|
Event Date 08/01/2016 |
Event Type
Injury
|
Manufacturer Narrative
|
As reported by the legal brief, the plaintiff is a citizen and resident of (b)(6).The plaintiff underwent placement of defendants' optease vena cava filter on or about (b)(6) 2014.The filter subsequently malfunctioned and caused injury and damages to the plaintiff, including, but not limited to, blood clots, clotting and occlusion of the inferior vena cava (ivc) filter, and recurrent deep vein thrombosis (dvt).As a direct and proximate result of these malfunctions, the plaintiff suffered life-threatening injuries and damages, and required extensive medical care and treatment.As a further proximate result, the plaintiff has suffered and will continue to suffer significant medical expenses, and pain and suffering, and other damages.The product was not returned for analysis.Additionally, as the sterile lot number was not available, device history record (dhr) review could not be performed.The optease is indicated for use in the prevention of recurrent pulmonary embolism (pe) via percutaneous placement in the vena cava for patients in which anticoagulants are contraindicated, anticoagulant therapy for thromboembolic disease has failed, emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced or for chronic, recurrent pulmonary embolism where anticoagulant therapy has failed, or is contraindicated.The purpose of a vena cava filter is to catch thrombus from the lower extremities as it travels along normal blood flow patterns up towards the heart.Neither blood clots, clotting nor thrombus within the vessel (or in the filter) represent a device malfunction.Rather, patient and pharmacological factors may have contributed to these events.The product¿s instructions for use (ifu) indicate that filter obstruction, blood clots and thrombus formation are potential complications of filter implantation.Given the limited information available for review at this time, there is nothing to suggest that the reported event is related to the design and manufacturing process of the device; therefore no corrective action will be taken.Should additional information become available, the file will be updated accordingly.
|
|
Event Description
|
As reported by the legal brief, the plaintiff is a citizen and resident of (b)(6).The plaintiff underwent placement of defendants' optease vena cava filter on or about (b)(6) 2014.The filter subsequently malfunctioned and caused injury and damages to the plaintiff, including, but not limited to, blood clots, clotting and occlusion of the inferior vena cava (ivc) filter, and recurrent deep vein thrombosis (dvt).As a direct and proximate result of these malfunctions, the plaintiff suffered life-threatening injuries and damages, and required extensive medical care and treatment.As a further proximate result, the plaintiff has suffered and will continue to suffer significant medical expenses, and pain and suffering, and other damages.
|
|
Manufacturer Narrative
|
The following additional information received per the patient profile from (ppf) indicates that the filter was fractured and that the ivc was occluded.The patient also reports to be suffering from anxiety.According to the medical records, the patient had a history of pulmonary embolism (pe) and right popliteal venous thrombosis and was not a candidate for long term anticoagulation.The trapease filter was deployed in the ivc below the renal veins and above the ivc bifurcation.The patient tolerated the procedure well and there were no reported complications.Additional information is pending and will be submitted within 30 days upon receipt.
|
|
Manufacturer Narrative
|
As reported, the patient underwent placement of an trapease inferior vena cava (ivc) filter.Per the medical records, the patient had a history of pulmonary embolism (pe) and right popliteal venous thrombosis and was not a candidate for long term anticoagulation.The trapease filter was deployed in the ivc below the renal veins and above the ivc bifurcation.The patient tolerated the procedure well and there were no reported complications.The filter subsequently malfunctioned and caused injury and damages to the patient, including, but not limited to, blood clots, clotting and occlusion of the inferior vena cava (ivc) filter, and recurrent deep vein thrombosis (dvt).As a direct and proximate result of these malfunctions, the patient suffered life-threatening injuries and damages, and required extensive medical care and treatment.As a further proximate result, the patient has suffered and will continue to suffer significant medical expenses, and pain and suffering, and other damages.Per the patient profile from (ppf), the filter was fractured and that the ivc was occluded.The patient also reports to be suffering from anxiety.The product was not returned for analysis as it remains implanted and the sterile lot number has not been provided; therefore, neither a device analysis nor a device history record review could be performed.The trapease vena cava filter is indicated for use in the prevention of recurrent pulmonary embolism (pe) via percutaneous placement in the vena cava for patients in which anticoagulants are contraindicated, anticoagulant therapy for thromboembolic disease has failed, emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced or for chronic, recurrent pulmonary embolism where anticoagulant therapy has failed, or is contraindicated.The purpose of a vena cava filter is to catch thrombus from the lower extremities as it travels along normal blood flow patterns up towards the heart.The instructions for use (ifu) states filter fracture is a potential complication of vena cava filters.Anatomic locations that create concentrated stress points from filter deformation (for example, deployment at apex of scoliosis, overlapping of either of the renal ostia, or placement adjacent to a vertebral osteophyte) may contribute to fracture of a particular filter strut.Blood clots and occlusive thrombosis within the filter do not represent a device malfunction.Anxiety does not represent a device malfunction and may be related to underlying patient related issues.Clinical factors that may have influenced the event include patient, pharmacological and lesion characteristics.Without procedural films or images for review the reported event(s) could not be confirmed.Given the limited information available for review at this time, there is nothing to suggest that the reported events are related to the design and manufacturing process of the device; therefore, no corrective action will be taken.
|
|