MAUDE Adverse Event Report: CORDIS CASHEL TRAPEASE PVCF FEM/JUG 55CM CSI; CATHETER, INTRAVASCULAR, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CORDIS CASHEL TRAPEASE PVCF FEM/JUG 55CM CSI; CATHETER, INTRAVASCULAR, DIAGNOSTIC Back to Search Results
Model Number 466FXXXX
Device Problems Occlusion Within Device (1423); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Occlusion (1984); Thrombosis (2100)
Event Date 08/01/2016
Event Type  Injury  
Manufacturer Narrative
As reported by the legal brief, the plaintiff is a citizen and resident of (b)(6).The plaintiff underwent placement of defendants' optease vena cava filter on or about (b)(6) 2014.The filter subsequently malfunctioned and caused injury and damages to the plaintiff, including, but not limited to, blood clots, clotting and occlusion of the inferior vena cava (ivc) filter, and recurrent deep vein thrombosis (dvt).As a direct and proximate result of these malfunctions, the plaintiff suffered life-threatening injuries and damages, and required extensive medical care and treatment.As a further proximate result, the plaintiff has suffered and will continue to suffer significant medical expenses, and pain and suffering, and other damages.The product was not returned for analysis.Additionally, as the sterile lot number was not available, device history record (dhr) review could not be performed.The optease is indicated for use in the prevention of recurrent pulmonary embolism (pe) via percutaneous placement in the vena cava for patients in which anticoagulants are contraindicated, anticoagulant therapy for thromboembolic disease has failed, emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced or for chronic, recurrent pulmonary embolism where anticoagulant therapy has failed, or is contraindicated.The purpose of a vena cava filter is to catch thrombus from the lower extremities as it travels along normal blood flow patterns up towards the heart.Neither blood clots, clotting nor thrombus within the vessel (or in the filter) represent a device malfunction.Rather, patient and pharmacological factors may have contributed to these events.The product¿s instructions for use (ifu) indicate that filter obstruction, blood clots and thrombus formation are potential complications of filter implantation.Given the limited information available for review at this time, there is nothing to suggest that the reported event is related to the design and manufacturing process of the device; therefore no corrective action will be taken.Should additional information become available, the file will be updated accordingly.
 
Event Description
As reported by the legal brief, the plaintiff is a citizen and resident of (b)(6).The plaintiff underwent placement of defendants' optease vena cava filter on or about (b)(6) 2014.The filter subsequently malfunctioned and caused injury and damages to the plaintiff, including, but not limited to, blood clots, clotting and occlusion of the inferior vena cava (ivc) filter, and recurrent deep vein thrombosis (dvt).As a direct and proximate result of these malfunctions, the plaintiff suffered life-threatening injuries and damages, and required extensive medical care and treatment.As a further proximate result, the plaintiff has suffered and will continue to suffer significant medical expenses, and pain and suffering, and other damages.
 
Manufacturer Narrative
The following additional information received per the patient profile from (ppf) indicates that the filter was fractured and that the ivc was occluded.The patient also reports to be suffering from anxiety.According to the medical records, the patient had a history of pulmonary embolism (pe) and right popliteal venous thrombosis and was not a candidate for long term anticoagulation.The trapease filter was deployed in the ivc below the renal veins and above the ivc bifurcation.The patient tolerated the procedure well and there were no reported complications.Additional information is pending and will be submitted within 30 days upon receipt.
 
Manufacturer Narrative
As reported, the patient underwent placement of an trapease inferior vena cava (ivc) filter.Per the medical records, the patient had a history of pulmonary embolism (pe) and right popliteal venous thrombosis and was not a candidate for long term anticoagulation.The trapease filter was deployed in the ivc below the renal veins and above the ivc bifurcation.The patient tolerated the procedure well and there were no reported complications.The filter subsequently malfunctioned and caused injury and damages to the patient, including, but not limited to, blood clots, clotting and occlusion of the inferior vena cava (ivc) filter, and recurrent deep vein thrombosis (dvt).As a direct and proximate result of these malfunctions, the patient suffered life-threatening injuries and damages, and required extensive medical care and treatment.As a further proximate result, the patient has suffered and will continue to suffer significant medical expenses, and pain and suffering, and other damages.Per the patient profile from (ppf), the filter was fractured and that the ivc was occluded.The patient also reports to be suffering from anxiety.The product was not returned for analysis as it remains implanted and the sterile lot number has not been provided; therefore, neither a device analysis nor a device history record review could be performed.The trapease vena cava filter is indicated for use in the prevention of recurrent pulmonary embolism (pe) via percutaneous placement in the vena cava for patients in which anticoagulants are contraindicated, anticoagulant therapy for thromboembolic disease has failed, emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced or for chronic, recurrent pulmonary embolism where anticoagulant therapy has failed, or is contraindicated.The purpose of a vena cava filter is to catch thrombus from the lower extremities as it travels along normal blood flow patterns up towards the heart.The instructions for use (ifu) states filter fracture is a potential complication of vena cava filters.Anatomic locations that create concentrated stress points from filter deformation (for example, deployment at apex of scoliosis, overlapping of either of the renal ostia, or placement adjacent to a vertebral osteophyte) may contribute to fracture of a particular filter strut.Blood clots and occlusive thrombosis within the filter do not represent a device malfunction.Anxiety does not represent a device malfunction and may be related to underlying patient related issues.Clinical factors that may have influenced the event include patient, pharmacological and lesion characteristics.Without procedural films or images for review the reported event(s) could not be confirmed.Given the limited information available for review at this time, there is nothing to suggest that the reported events are related to the design and manufacturing process of the device; therefore, no corrective action will be taken.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRAPEASE PVCF FEM/JUG 55CM CSI
Type of Device
CATHETER, INTRAVASCULAR, DIAGNOSTIC
Manufacturer (Section D)
CORDIS CASHEL
cahir road
cashel, co. tipperary
miami FL
EI 
MDR Report Key6515432
MDR Text Key73493903
Report Number1016427-2017-00285
Device Sequence Number1
Product Code DQO
Combination Product (y/n)N
PMA/PMN Number
K020316
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 01/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/24/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number466FXXXX
Device Catalogue Number466P306AU
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Distributor Facility Aware Date03/29/2017
Date Manufacturer Received01/09/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Life Threatening;
Patient Age89 YR
-
-