Correction: initial mdr was submitted with awareness date of 04/06/2016.The correct awareness date should be 04/06/2017.Integra has completed their internal investigation on may 16, 2017.Results: evaluation of returned device; no issues observed.Unit cleaned per protocol 1107.The unit was setup per the ifu, then the part passed the1101 testing.With respect to the returned unit it has passed all specific functional testing requirements with the following exception: the lock had rotational movement when the unit was not under pressure.Dhr review; no abnormalities related to the reported failure.Complaints history; no manufacturing or design related trend has been identified.Post market information will continue to be monitored.Conclusion: the reported complaint was not confirmed.The unit was setup per the instructions for use and the part passed the1101 inspection testing.With respect to the returned unit it has passed all specific functional testing requirements with the following exception: the lock had rotational movement when the unit was not under pressure.The rotational movement would not have caused/contributed to the reported harm of slippage.When the unit is properly positioned and put under pressure, the unit will not slip.
|