• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAKO SURGICAL CORP. TIBIAL INLAY IMPACTOR HEAD; PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAKO SURGICAL CORP. TIBIAL INLAY IMPACTOR HEAD; PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM Back to Search Results
Catalog Number 160177
Device Problems Break (1069); Fracture (1260)
Patient Problems No Consequences Or Impact To Patient (2199); No Information (3190)
Event Date 03/28/2017
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of the investigation.
 
Event Description
The impactor tip part number 160177 broke at the hospital.
 
Manufacturer Narrative
An event regarding crack/fracture involving an mako impactor was reported.The event was confirmed.Method & results: device evaluation and results: the device was returned in used condition and showed the impactor head was fractured.There are scratches and gouges on the bottom and sides of the impactor.Material analysis indicated fracture consistent with an overload condition.Medical records received and evaluation: not performed as medical records were not received and patient factors didn't contribute to the event.Device history review: indicated the devices accepted into final stock from the reported lot were free from discrepancies.Complaint history review: there has been no other event for the lot referenced.Conclusions: the reported event was confirmed by visual inspection which showed the impactor head fractured.Material analysis indicated fracture consistent with an overload condition.No further investigation for this event is possible at this time.If additional information will be received, this investigation will be reopened and re-evaluated.
 
Event Description
The impactor tip part number 160177 broke at the hospital.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TIBIAL INLAY IMPACTOR HEAD
Type of Device
PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM
Manufacturer (Section D)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer (Section G)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer Contact
cindy chuhinko
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key6517188
MDR Text Key73735140
Report Number3005738134-2017-00004
Device Sequence Number1
Product Code NPJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K090763
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/26/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/24/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number160177
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/17/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received06/29/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-