Additional narrative/investigation ¿ evaluation: a review of the following investigation methods were done on the returned device: functional test, review of complaint history, review of device history record, review of manufacturing instructions, review of quality control, and visual inspection.One stone extractor was returned for investigation.The device was returned with the handle in the closed position and the basket formation was in the open position.The collet knob was noted to be loose and partially unscrewed.There is a gap between the knob and the handle.The support sheath is adhered securely to the basket sheath.A visual examination of the device noted the basket sheath has a kink from the end of the support sheath.A functional test was performed and the unidex handle (udh) would not actuate the basket formation.The collet knob was tightened for investigational purposes and the udh handle was found to actuate the basket formation several times.The device history record was reviewed and no non-conformances were noted for reported lot number.Based on the provided information a definitive root cause cannot be established or reported at this time.We will notify the appropriate personnel and continue to monitor for similar complaints.Measures have been previously initiated to address this issue.Quality engineering risk assessment (qera) was revised to address this issue.No further action is required.
|