• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN KANGAROO; TUBES, GASTROINTESTINAL (AND ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN KANGAROO; TUBES, GASTROINTESTINAL (AND ACCESSORIES) Back to Search Results
Model Number 383400
Device Problem Thermal Decomposition of Device (1071)
Patient Problem No Patient Involvement (2645)
Event Date 03/30/2017
Event Type  malfunction  
Manufacturer Narrative
Submit date: 4/25/2017.
 
Event Description
It was reported to covidien that a customer had an issue with an enteral feeding pump.Upon triage on (b)(6) 2017, service found that the unit had a ruptured component from potential overheating.
 
Manufacturer Narrative
An evaluation of the kangaroo pump was performed for the reported condition of, ¿the customer reports the unit is not working.Upon triage, the service tech found the unit had a ruptured component from potential overheating.¿.The unit was triaged and the reported issue was confirmed.The potential root causes are the use of an unauthorized power adapter by the user and a possibly defective component.All device history records are reviewed for quality inspections and parameter compliance prior to releasing the product for shipment.Complaint trending information is being reviewed on a monthly basis and if a trend is observed, actions will be taken as necessary.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KANGAROO
Type of Device
TUBES, GASTROINTESTINAL (AND ACCESSORIES)
Manufacturer (Section D)
COVIDIEN
15 hampshire street
mansfield MA 02048
Manufacturer (Section G)
COVIDIEN
15 hampshire street
mansfield MA 02048
Manufacturer Contact
edward almeida
15 hampshire street
mansfield, MA 02048
5084524151
MDR Report Key6519113
MDR Text Key73612194
Report Number1282497-2017-05104
Device Sequence Number1
Product Code LZH
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/06/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number383400
Device Catalogue Number383400
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/20/2017
Initial Date FDA Received04/25/2017
Supplement Dates Manufacturer Received06/07/2017
Supplement Dates FDA Received11/06/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-