• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. ENDOSCPC CANN.DRL.BIT 4.5 STRL; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. ENDOSCPC CANN.DRL.BIT 4.5 STRL; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT Back to Search Results
Catalog Number 7207315
Device Problem Sticking (1597)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/18/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(6).
 
Event Description
It was reported the device was stuck in the femoral tunnel during surgery.
 
Manufacturer Narrative
Examination was not possible, as the device was not returned.The investigation was limited to the information provided.The investigation could not draw any conclusions about the reported event with the clinical details provided.A review of the device history record was performed which confirmed no inconsistencies.No further investigation is warranted at this time.
 
Manufacturer Narrative
A visual assessment of the device confirmed the reported breakage.The entire drill head has broken from the shaft.The broken pieces were not returned for evaluation.The drill shaft is bent.The condition of the device indicates it was being used over a bent guidewire which would require the use of excessive force.Per the device ifu ¿as with any surgical instrument, careful attention should be made to assure that excessive force is not placed on the instrument.Excessive forces applied to the instrument can result in failure¿.Further investigation is not warranted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOSCPC CANN.DRL.BIT 4.5 STRL
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048
Manufacturer Contact
jim gonzales
7000 w. william cannon
austin, TX 78735
MDR Report Key6520940
MDR Text Key73810858
Report Number1219602-2017-00426
Device Sequence Number1
Product Code HWE
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 07/14/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/25/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number7207315
Device Lot Number50560405
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/07/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received04/04/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-