(b)(4).Any additional information received from the customer will be included in a follow-up report.(b)(4).Results of investigation: the carefusion manufacturing service group received the suspect mark 7 respirator device for investigation.The mark 7 was visually and physically inspected, which found the pressure manometer reading 15 cmh2o on the device.The reported event of the device failing to cycle off was not duplicated, therefore no component failure investigation was performed and no root cause can be determined.The manometer gauge failure was confirmed and a replacement pressure manometer was installed on the device.The device was tested, which met manufacture specifications, so no further investigation will be required.
|