• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HILL-ROM BATESVILLE VERSACARE FRAME; A/C POWERED ADJUSTABLE HOSPITAL BED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HILL-ROM BATESVILLE VERSACARE FRAME; A/C POWERED ADJUSTABLE HOSPITAL BED Back to Search Results
Model Number P3200K000004
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/19/2017
Event Type  malfunction  
Manufacturer Narrative
The hill-rom technician found the external alarm needed to be replaced.The piezo alarm is the source of the brake not set alarm.Per the hill-rom service manual, perform annual preventive maintenance procedures to make sure all versacare® bed components are functioning as originally designed.Brake casters should be checked for cuts, wear and quality of tread, etc.And replaced when necessary.Examine the brakes to see whether the bed moves when the brake are set.Replace as necessary.A search of the hill-rom maintenance records showed hill-rom performed preventative maintenance on this bed in 2017.It is unknown if the facility performed any other preventative maintenance on this bed.The technician replaced the external alarm to resolve the issue.Based on this information, no further action is required.
 
Event Description
Hill-rom received a report from a hill-rom technician stating the bed alarm was not working.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint #(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSACARE FRAME
Type of Device
A/C POWERED ADJUSTABLE HOSPITAL BED
Manufacturer (Section D)
HILL-ROM BATESVILLE
1069 state route 46 east
batesville IN 47006
Manufacturer Contact
erin padgett
1069 state route 46 east
batesville, IN 47006
8129312791
MDR Report Key6540120
MDR Text Key74412458
Report Number1824206-2017-00203
Device Sequence Number1
Product Code FNL
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 04/19/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/03/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other Caregivers
Device Model NumberP3200K000004
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received04/19/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/13/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-