Model Number 37602
Device Problems Electromagnetic Interference (1194); Failure to Interrogate (1332); Device Displays Incorrect Message (2591); Communication or Transmission Problem (2896)
Patient Problem Therapeutic Effects, Unexpected (2099)
Event Type  Injury  
Event Description
A manufacturer representative (rep) reported that the patient's therapy has stopped, with a related power on reset (por) that occurred.The rep is to meet with the patient on (b)(6) regarding the issue.No further complications are anticipated.The patient's indication for implant is unknown.
 
Event Description
Additional information was received by the rep who reported the therapy stoppage occurred after the patient was defibrillated for reasons unrelated to the device.They could no longer communicate with the device and it was reported that it would be replaced the day after this secondary report.No further complications are anticipated.The patient's indications for use were dystonia and movement disorders.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIVA
Type of Device
IMPLANTED SUBCORTICAL ELECTRICAL STIMULATOR (MOTOR DISORDERS)
Manufacturer (Section D)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
road 31, km. 24, hm 4
ceiba norte industrial park
juncos,pr MN 00777 1200
Manufacturer (Section G)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
road 31, km. 24, hm 4
ceiba norte industrial park
juncos,pr MN 00777 1200
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key6541429
MDR Text Key74320925
Report Number3007566237-2017-01708
Device Sequence Number1
Product Code MRU
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
H020007
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date 05/22/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/28/2013
Device Model Number37602
Device Catalogue Number37602
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 05/01/2017
Initial Date FDA Received05/03/2017
Supplement Dates Manufacturer ReceivedNot provided
05/03/2017
05/03/2017
Supplement Dates FDA Received05/22/2017
10/03/2017
10/04/2017
Date Device Manufactured07/11/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-