• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER MEDICAL-KALAMAZOO GOBED II; BED, AC-POWERED ADJUSTABLE HOSPITAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER MEDICAL-KALAMAZOO GOBED II; BED, AC-POWERED ADJUSTABLE HOSPITAL Back to Search Results
Catalog Number FL28EX
Device Problems Thermal Decomposition of Device (1071); Component Missing (2306)
Patient Problem Electric Shock (2554)
Event Date 04/05/2017
Event Type  Injury  
Event Description
It was alleged that a nurse experienced a shock when plugging a power cord into the auxiliary outlet on the bed.Upon inspection by the service technician, it was found that the outlet did have burn marks, and that the outlet had a missing screw.The nurse reportedly went to the emergency room for evaluation, but the results of the evaluation were not reported, therefore the severity of the alleged shock and if any medical intervention was required is not known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GOBED II
Type of Device
BED, AC-POWERED ADJUSTABLE HOSPITAL
Manufacturer (Section D)
STRYKER MEDICAL-KALAMAZOO
3800 east centre avenue
portage MI 49002
Manufacturer (Section G)
STRYKER MEDICAL-KALAMAZOO
3800 east centre avenue
portage MI 49002
Manufacturer Contact
mary klaver
3800 east centre avenue
portage, MI 49002
2693292100
MDR Report Key6544175
MDR Text Key74413198
Report Number0001831750-2017-00174
Device Sequence Number1
Product Code FNL
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 05/04/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue NumberFL28EX
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 04/05/2017
Initial Date FDA Received05/04/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/24/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-