• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AGA MEDICAL CORPORATION AMPLATZER SEPTAL OCCLUDER; CARDIAC OCCLUSION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AGA MEDICAL CORPORATION AMPLATZER SEPTAL OCCLUDER; CARDIAC OCCLUSION DEVICE Back to Search Results
Model Number 9-ASD-022
Device Problem Retraction Problem (1536)
Patient Problem No Code Available (3191)
Event Date 04/08/2017
Event Type  Injury  
Event Description
The right atrial disc of a 22 mm amplatzer septal occluder (aso) deformed upon deployment.The aso was not able to be retracted back into the 10f cook mullins sheath and the user was forced to release the aso.The patient was sent to surgery for aso retrieval and surgical closure of the defect.The patient is reported to be stable.
 
Manufacturer Narrative
The results of this investigation confirmed the amplatzer septal occluder met all functional specifications when analyzed at abbott.A review of the device history record confirmed the occluder met all visual, dimensional, and functional specifications at the time it was manufactured, prior to shipment.There was no evidence to suggest there was an intrinsic defect in the occluder, and the cause for the reported event remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER SEPTAL OCCLUDER
Type of Device
CARDIAC OCCLUSION DEVICE
Manufacturer (Section D)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key6544433
MDR Text Key74421649
Report Number2135147-2017-00055
Device Sequence Number1
Product Code MLV
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
P000039
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/22/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/04/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date03/31/2021
Device Model Number9-ASD-022
Device Catalogue Number9-ASD-022
Device Lot Number5489566
Other Device ID Number00811806010182
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/16/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/19/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/29/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age6 YR
-
-