MAUDE Adverse Event Report: GE MEDICAL SYSTEMS (CHINA) CO., LTD (WUXI) CARESTATION 650; ANESTHESIA GAS MACHINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GE MEDICAL SYSTEMS (CHINA) CO., LTD (WUXI) CARESTATION 650; ANESTHESIA GAS MACHINE Back to Search Results
Device Problems Loss of Power (1475); Failure to Power Up (1476); Computer Operating System Problem (2898); Device Operates Differently Than Expected (2913)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 04/06/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Ge healthcare¿s investigation into the reported occurrence is ongoing.A follow-up report will be issued when the investigation has been completed.Device evaluation anticipated, but not yet begun.
 
Event Description
The hospital reported that, there was a complete shutdown of the machine and all ventilation was lost.There was a continuous beep heard from the back of the machine until a power on/off cycle was attempted.Rebooting failed to turn the machine on, the display did not power on, although the beep did stop.The machine was swapped with another to continue the procedure.
 
Manufacturer Narrative
A ge healthcare service representative replaced the complete display assembly containing the central processor unit printed circuit board, calibrated and configured the machine, and upgraded the system software from sp3 to sp4.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARESTATION 650
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
GE MEDICAL SYSTEMS (CHINA) CO., LTD (WUXI)
no. 19 changjiang road
national hi-tech dev. zone
wuxi
CH 
Manufacturer Contact
joseph seliga
3000 n grandview blvd.
waukesha, WI 
MDR Report Key6547361
MDR Text Key74633088
Report Number9710602-2017-00087
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
PMA/PMN Number
K151570
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup
Report Date 05/05/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Other Device ID NumberSEE H10
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/06/2017
Initial Date FDA Received05/05/2017
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received05/16/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/01/1970
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N