Device Problems
Loss of Power (1475); Failure to Power Up (1476); Computer Operating System Problem (2898); Device Operates Differently Than Expected (2913)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
Event Date 04/06/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Ge healthcare¿s investigation into the reported occurrence is ongoing.A follow-up report will be issued when the investigation has been completed.Device evaluation anticipated, but not yet begun.
|
|
Event Description
|
The hospital reported that, there was a complete shutdown of the machine and all ventilation was lost.There was a continuous beep heard from the back of the machine until a power on/off cycle was attempted.Rebooting failed to turn the machine on, the display did not power on, although the beep did stop.The machine was swapped with another to continue the procedure.
|
|
Manufacturer Narrative
|
A ge healthcare service representative replaced the complete display assembly containing the central processor unit printed circuit board, calibrated and configured the machine, and upgraded the system software from sp3 to sp4.
|
|