Reportable based on device analysis completed on 19-april-2017.It was reported that electrogram interference occurred.A blazer prime¿ xp was selected for use.During the latter stage of the procedure, and following a number of ablations to the cavotricuspid isthmus (cti), an intermittent loss of the electrogram was noted,when the ablation distal electrogram became raised of the baseline shown on the recording system.This occurred intermittently without any noticeable change in the catheter position.The ablation proximal electrogram was unaffected.The procedure was completed with another of the same device.No patient complications reported and patient's status was stable.However, device analysis revealed that the r2 and r3 seals were compromised; there is body fluid along the edge of the rings.
|