Reportable based on device analysis completed on 26-apr-2017.It was reported that shaft kink occurred.The 90% stenosed target lesion was located in a moderately tortuous and severely calcified coronary artery.A 12 x 2.50 promus premier¿ drug-eluting stent was advanced to treat the lesion.However, during the procedure, the stent delivery system shaft was kinked.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.However, returned device analysis revealed stent damage.
|