MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH HF-RESECTION ELECTRODE, LOOP, 24 FR., 0.35 WIRE, 12°; HF-RESECTION ELECTRODES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH HF-RESECTION ELECTRODE, LOOP, 24 FR., 0.35 WIRE, 12°; HF-RESECTION ELECTRODES Back to Search Results
Model Number A22201A
Device Problems Break (1069); Detachment Of Device Component (1104); Nonstandard Device (1420)
Patient Problem Foreign Body In Patient (2687)
Event Date 04/24/2017
Event Type  malfunction  
Manufacturer Narrative
The suspect medical device was not returned to olympus for evaluation/investigation since it was reportedly discarded by the user facility after the procedure.Therefore, the exact cause of the user's experience and the reported phenomenon could not be determined and is being judged as unknown.However, olympus america inc.("oai") is implementing a removal action of specified lot numbers of the monopolar hf resection electrodes a22201c and wa22037c.The electrodes are used for endoscopic diagnosis and treatment in urological and gynecological applications.Oai has initiated this removal action after receiving an increased number of complaints regarding loop wires breaking at the distal end of the referenced electrodes.Investigations have confirmed that loop wires can break during the intended use of the electrodes.As a result, a fragment may fall inside the patient and will need to be retrieved.Retrieval of the fragment could prolong the procedure and, under certain circumstances, could require additional surgical treatment.The investigation revealed that the loop wires of the affected electrodes were damaged during production.The cause of this damage is defective manufacturing equipment.The damaged loop wires cannot be detected by visual inspection.There has been no report to date of an adverse event or patient injury.However, in an effort to prevent a potential risk to patient health, oai is undertaking this action to remove the affected lot numbers.Oai's correction number according to 21 cfr 806.10 (c) (1): 2429304-4/18/2017-044r.
 
Event Description
Olympus was informed that during a therapeutic transurethral resection (tur) procedure, the loop wire at the distal end of the hf resection electrode broke off and fell inside the patient.However, no fragments remained inside the patient since they were reportedly retrieved.The intended procedure was successfully completed with another hf resection electrode and there was no adverse event or patient injury.This is report 2 out of 3.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HF-RESECTION ELECTRODE, LOOP, 24 FR., 0.35 WIRE, 12°
Type of Device
HF-RESECTION ELECTRODES
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS MEDICAL PRODUCTS CZECH, SPOL. S R.O.
telickova 457/29
prerov 75124
EZ   75124
Manufacturer Contact
daniel wladow
kuehnstrasse 61
hamburg 22045
GM   22045
4940669662
MDR Report Key6553530
MDR Text Key74899976
Report Number9610773-2017-00064
Device Sequence Number1
Product Code GCP
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
PK790071
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Nurse
Remedial Action Recall
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/09/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberA22201A
Device Catalogue NumberA22201A
Device Lot NumberP16X0001
Other Device ID Number04042761021302
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/24/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/01/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-