Model Number 19AGFN-756
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Thrombus (2101)
Event Date 04/11/2017
Event Type  Injury  
Event Description
On (b)(6) 2017, a 19mm regent mechanical heart valve was implanted in the aortic position.On (b)(6) 2017, the valve was explanted secondary to reports the valve was not operating as intended due to a suspected thrombus.An 18mm medtronic ats valve was implanted.The patient was reported to be doing well and recovering.
 
Manufacturer Narrative
The results of this investigation concluded there was inflow and outflow organizing fibrin thrombus in all four of the recessed pivot areas.Both leaflets were immobile when manually manipulated.No acute inflammation was present in the valve.A review of the device history record showed the device met specifications prior to leaving abbott manufacturing facilities.There was no evidence found to suggest the cause of the thrombus was due to an intrinsic defect in the valve, as supported by review of the valve's device history record and the analysis performed.The cause of the reported event remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SJM REGENT HEART VALVE W/FLEX CUFF
Type of Device
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ST. JUDE MEDICAL PUERTO RICO, INC.
p.o. box 998
lot 20 b st.
caguas, puerto rico 00725
Manufacturer (Section G)
ST. JUDE MEDICAL PUERTO RICO, INC.
p.o. box 998
lot 20 b st.
caguas, puerto rico 00725
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key6554079
MDR Text Key74739327
Report Number2648612-2017-00055
Device Sequence Number1
Product Code LWQ
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
P810002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/18/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/09/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date01/18/2022
Device Model Number19AGFN-756
Device Catalogue Number19AGFN-756
Device Lot Number5818719
Other Device ID Number05414734005852
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/01/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/16/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/19/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age6 MO
-
-