Model Number CD3371-40QC
Device Problems Pacing Inadequately (1442); Device Displays Incorrect Message (2591); Programming Issue (3014)
Patient Problem Therapeutic Response, Decreased (2271)
Event Date 03/30/2017
Event Type  malfunction  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
Upon interrogation, the right ventricular lead threshold was noted to be increased following suspected micro-dislodgement of the lead during a mitraclip procedure.During a manual threshold test, a parameter conflict error appeared and the lead switched to an rv-only pacing function instead of the required biventricular pacing.The patient was put on temporary external pacing since the rv lead could not effectively pace the patient.The device had to be reprogrammed to nominal settings to resume normal pacing possibly due to the higher threshold values because of the dislodgement.The parameter error was attributed to possible issues with telemetry during interrogation.The patient is recovering.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA MP ICD
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
645 almanor avenue
sunnyvale CA 94085
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
645 almanor avenue
sunnyvale CA 94085
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key6558687
MDR Text Key74869806
Report Number2938836-2017-23052
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 05/10/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/10/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2018
Device Model NumberCD3371-40QC
Device Catalogue NumberCD3371-40QC
Device Lot NumberA000018115
Other Device ID Number05414734508407
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/30/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/13/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-