• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SPECTRANETICS SPECTRANETICS LEAD LOCKING DEVICE; LLD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SPECTRANETICS SPECTRANETICS LEAD LOCKING DEVICE; LLD Back to Search Results
Model Number UNAVAILABLE
Device Problem Break (1069)
Patient Problem No Information (3190)
Event Date 04/12/2017
Event Type  Injury  
Manufacturer Narrative
Device model number, lot number, udi number, and expiration date are unavailable.The device was discarded by the user before this information could be obtained.Device manufacture date is dependent on device lot number, thus is unavailable.
 
Event Description
It was reported that during a lead extraction procedure to remove two non-functional cardiac leads (ra and rv), the rv lead broke.Reportedly, the rv lead had been prepped with a spectranetics lead locking device (lld) and a glidelight laser sheath and tightrail device were used to make progress down the lead.When traction force was applied to the lead, it came apart.The remainder of the lead was reportedly retrieved from the patient using femoral snares.The procedure was completed successfully and both leads were removed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPECTRANETICS LEAD LOCKING DEVICE
Type of Device
LLD
Manufacturer (Section D)
SPECTRANETICS
9965 federal drive
colorado springs CO 80921
Manufacturer (Section G)
SPECTRANETICS
Manufacturer Contact
nick mcnabb
9965 federal drive
colorado springs, CO 80921
719447-257
MDR Report Key6559855
MDR Text Key74934413
Report Number1721279-2017-00079
Device Sequence Number1
Product Code DRB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K043401
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Physician
Type of Report Initial
Report Date 04/17/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Model NumberUNAVAILABLE
Device Catalogue NumberUNAVAILABLE
Device Lot NumberUNAVAILABLE
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/17/2017
Initial Date FDA Received05/11/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
BOSTON SCIENTIFIC ICD LEAD 4243 (IMPL 215MO); BOSTON SCIENTIFIC PACING LEAD 4034 (IMPL 215MO); SPECTRANETICS CVX-300 EXCIMER LASER SYSTEM; SPECTRANETICS GLIDELIGHT LASER SHEATH; SPECTRANETICS LEAD LOCKING DEVICE; SPECTRANETICS TIGHTRAIL ROTATING DILATOR SHEATH
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
Patient Age71 YR
Patient Weight59
-
-