MAUDE Adverse Event Report: BOSTON SCIENTIFIC - GALWAY CAROTID WALLSTENT®; STENT, CAROTID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - GALWAY CAROTID WALLSTENT®; STENT, CAROTID Back to Search Results
Model Number M001719050
Device Problem Migration or Expulsion of Device (1395)
Patient Problem Device Embedded In Tissue or Plaque (3165)
Event Type  Injury  
Manufacturer Narrative
Age at time of event: 18 years or older.(b)(4).If implanted, give date: (b)(6) 2017.Device evaluated by mfr: the complaint device was not returned for analysis.The batch number is unknown and the manufacturing records for the complaint device could not be reviewed.The root cause is anticipated procedural complications as this event is a known physiological effect of the procedure and is noted within the dfu.(b)(4).
 
Event Description
It was reported that stent migration occurred.In (b)(6) 2017, a 10 x 31,5.9f,135 cm carotid wallstent® was deployed in the carotid artery.In (b)(6) 2017, during a follow up visit, it was noted that the carotid wallstent® had migrated into the internal carotid artery.The patient was scheduled for re-treatment.No further patient complications were reported and the patient's status was good.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAROTID WALLSTENT®
Type of Device
STENT, CAROTID
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
Manufacturer (Section G)
BOSTON SCIENTIFIC - GALWAY
Manufacturer Contact
sonali arangil
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key6573273
MDR Text Key75379529
Report Number2134265-2017-04767
Device Sequence Number1
Product Code NIM
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P050019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 04/26/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/17/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM001719050
Device Catalogue Number71-905
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/26/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-