DEPUY SYNTHES POWER TOOLS 8 CM HEAVY DUTY MEDIUM ATTACHMENT; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
|
Back to Search Results |
|
Catalog Number MEDIUM-HD-G1_LL |
Device Problem
Detachment Of Device Component (1104)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 04/26/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The actual device was returned for evaluation.Reliability engineering evaluated the device and observed that the device bearing stack screw was missing causing the bearings and the spacers to fall out of the device.It was also observed that the loctite failed causing the bearing stack screw to fall out.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to component damage from normal wear from use and servicing over time.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
It was reported that during pre-surgery setup, it was observed that the attachment device inner parts fell out.There were no delays to the planned surgical procedure.A spare device was available for use.There was no patient involvement reported.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|