Catalog Number A3059
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Investigation completed on 05/may2017: unit was per functional inspection procedures and no duplicate condition was found as reported by the customer.Some movement in the swivel was noted but it is unrelated to the reported complaint failure.Device history record reviewed for product id 40a3059 work order (b)(4) serial # (b)(4) a total of (b)(4) were manufactured on 10-20-16 show no abnormalities related to the reported failure.The devices manufactured during this period passed all required inspection points with no associated mrr¿s, variances or rework.No service history is on file for this device.A two year lookback in the complaint system for this reported failure and or related to " does not lock correctly or issue with lock" for product id a3059 shows that 7 complaints were received including this case.No new design or manufacturing trends have been identified.This issue will be monitored.Root cause was unable to be determined based on unit passing all required tests for final inspection.
 
Event Description
The doctor had trouble with the unit; he did not think it locked correctly.The doctor stated he would rather have integra check out the unit; even if integra does not find anything wrong.There was no patient injury.Additional information was requested but to date, no response was received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD COMPOSITE SERIES SKULL CLAMP
Type of Device
¿COMPOSITE SERIES¿
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH 45227
Manufacturer (Section G)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
cincinnati OH 45227
Manufacturer Contact
rowena bunuan
311 enterprise drive
plainsboro, NJ 08536
6099362393
MDR Report Key6576907
MDR Text Key75526717
Report Number3004608878-2017-00168
Device Sequence Number1
Product Code HBL
Combination Product (y/n)N
PMA/PMN Number
K142238
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 03/07/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/18/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberA3059
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/03/2017
Date Manufacturer Received05/05/2017
Date Device Manufactured10/20/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown