ENTEROMEDICS, INC. MAESTRO RECHARGEABLE SYSTEM; POSTERIOR LEAD Back to Search Results
Model Number 2200P-47E
Device Problems Device Inoperable (1663); Device Displays Incorrect Message (2591); Material Distortion (2977); Material Twisted/Bent (2981)
Patient Problems Therapeutic Response, Decreased (2271); No Known Impact Or Consequence To Patient (2692)
Event Date 04/24/2017
Event Type  malfunction  
Manufacturer Narrative
Explanted device not yet returned.
 
Event Description
This subject is a participant in the (b)(6) clinical trial.This subject was originally in the control arm of the trial and underwent implantation of the active maestro rechargeable system on (b)(6) 2014.Impedance issues were reported related to the implanted anterior lead (reference mdr3005025697-2017-00001 for information on that component).During a revision procedure to replace the anterior lead, it was discovered that the anterior lead and posterior lead were twisted around each other in the pocket of the rechargeable neuroregulator.As a result, the posterior lead was also replaced.The revision procedure was successfully completed without adverse impact to the subject.Explanted device components have not yet been received by enteromedics.Reference mdr3005025697-2017-00001 for report related to the anterior lead.
 
Event Description
This subject is a participant in the (b)(6) clinical trial.This subject was originally in the control arm of the trial and underwent implantation of the active maestro rechargeable system on (b)(6) 2014.Impedance issues were reported related to the implanted anterior lead (reference mdr 3005025697-2017-00001 for information on that component).During a revision procedure to replace the anterior lead, it was discovered that the anterior lead and posterior lead were twisted around each other in the pocket of the rechargeable neuroregulator.As a result, the posterior lead was also replaced.The revision procedure was successfully completed without adverse impact to the subject.Explanted device components have not yet been received by enteromedics.Reference mdr 3005025697-2017-00001 for report related to the anterior lead.Supplement 001: explanted device component received by enteromedics on june 1, 2017.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAESTRO RECHARGEABLE SYSTEM
Type of Device
POSTERIOR LEAD
Manufacturer (Section D)
ENTEROMEDICS, INC.
2800 patton road
saint paul MN 55113
Manufacturer (Section G)
ENTEROMEDICS, INC.
2800 patton road
saint paul MN 55113
Manufacturer Contact
randy hoyt
2800 patton road
saint paul, MN 55113
6517892671
MDR Report Key6589490
MDR Text Key75967249
Report Number3005025697-2017-00014
Device Sequence Number1
Product Code PIM
UDI-Device Identifier00857334004293
UDI-Public00857334004293
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P130019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,study
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/29/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date01/01/2017
Device Model Number2200P-47E
Device Catalogue Number2200P-47E
Device Lot NumberCR-01826
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/01/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/24/2017
Initial Date FDA Received05/24/2017
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received06/29/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age64 YR
Patient Weight114
-
-