• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION OCTRODE LEAD KIT, 60CM LENGTH; SCS LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION OCTRODE LEAD KIT, 60CM LENGTH; SCS LEAD Back to Search Results
Model Number 3186
Device Problems Device Stops Intermittently (1599); Low impedance (2285)
Patient Problems Fall (1848); Inadequate Pain Relief (2388)
Event Date 03/01/2017
Event Type  Injury  
Manufacturer Narrative
Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Event Description
Device 3 of 3: reference mfr report: 3006705815-2017-00578 & 1627487-2017-02803.It was reported the patient experienced a fall and the patient's scs stimulation was turning on and off since.A company representative met with the patient and troubleshooting of the patient¿s system revealed low impedance for all of the leads contacts.Follow up identified surgical intervention was taken, during the procedure the physician suspected fluid may have entered into the ipg header.In addition, the patient's ipg was inoperable, the physician replaced the ipg.Reportedly, stimulation therapy was restored postoperatively.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OCTRODE LEAD KIT, 60CM LENGTH
Type of Device
SCS LEAD
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
MDR Report Key6590507
MDR Text Key75962729
Report Number1627487-2017-02804
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial
Report Date 05/25/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/28/2016
Device Model Number3186
Device Lot Number4409876
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 05/01/2017
Initial Date FDA Received05/25/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/08/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
MODEL 1192 (X2), SCS ANCHOR
Patient Outcome(s) Other;
Patient Age64 YR
-
-