• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 ROCHESTER MEC WIDEBAND M 30BX

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 ROCHESTER MEC WIDEBAND M 30BX Back to Search Results
Model Number 36302
Device Problem Physical Property Issue (3008)
Patient Problem Pain (1994)
Event Date 05/05/2017
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event Description
It was reported that the condoms had too much adhesive, and caused slight pain during removal.
 
Manufacturer Narrative
Received 3 unopened wideband mecs.The reported event was unconfirmed since the samples met specification.The samples were visually inspected for holes or damage on the package, legibility of printed numbers and letters, expiration date, correct impression, print color, missing parts of printing, position of printing, damaged component, foreign matter.The samples were evaluated according to the alleged failure mode: "too much adhesive".The evaluation could not confirm the reported event since the samples were within specification, the acceptable peel strength must be in a range from 0.80 to 2.80lbf for a wideband mec and sample results were: sample 1: 1.47 lbf, sample 2: 1.78 lbf, sample 3: 2.06 lbf.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use state the following: "description/indication the self-adhering male external catheter is designed for the management of male urinary incontinence.Contraindication do not use on irritated or compromised skin.Precaution do not use if allergic reaction occurs.For good hygiene, change catheter daily.Use of a single device for longer periods than 24 hours may increase the risk of complications.Directions: to apply wash penis with mild soap and warm water.Dry thoroughly.Trim pubic hair if necessary.Open package at perforation.To remove plastic insert, squeeze catheter at the top of the white cone and pull to release.Unroll self-adhering catheter over penis.Gently squeeze the catheter to properly seal adhesive to the skin.Connect to collection bag.Important: wear time may be significantly reduced if adhesive is not properly sealed to the skin.Directions: to remove gently roll catheter off the penis.Note: if necessary, apply a warm wet compress (such as a wet washcloth) around the catheter to help loosen the adhesive".(b)(4).
 
Event Description
It was reported that the condoms had too much adhesive, and caused slight pain during removal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROCHESTER MEC WIDEBAND M 30BX
Type of Device
MEC
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
parque industrial aeropuerto
covington GA 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
parque industrial aeropuerto
covington GA 30014
Manufacturer Contact
amy gravley
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key6595451
MDR Text Key76237654
Report Number1018233-2017-02750
Device Sequence Number1
Product Code NNX
UDI-Device Identifier00801741071041
UDI-Public(01)00801741071041
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,other
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/07/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/26/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date03/31/2021
Device Model Number36302
Device Catalogue Number36302
Device Lot NumberNGAR4507
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/26/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received12/06/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/07/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-