Catalog Number LSMU1350626 |
Device Problems
Difficult to Remove (1528); Failure to Advance (2524)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 05/09/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.Receipt of device is pending.The investigation is currently in progress.
|
|
Event Description
|
It was reported that two devices could not be advanced through a 6f introducer sheath.There were no anomalies noted during the prep of the devices.There was no damage noted to the box, tray or stent cover and there was no difficulty removing the stent cover on either of the devices.The intended treatment site was the renals.The delivery system was flushed.Access was made via the iliac.A benson guidewire and a 6f ansel cook introducer sheath were used.The vessel was pre-dilated using a 6x4 armada device.Two vascular covered stent delivery systems allegedly did not go through a 6fr introducer sheath during an attempt to treat the renals via the iliacs.In both cases the device advanced about 2 inches into the sheath and then failed to advance further.The devices were removed with the sheath.There was difficulty retracting the devices from the patient as it was reported that it was a little tight.Multiple attempts to insert the devices were not made.Air evacuation was performed prior to inserting both devices.The delivery systems were not pulled back into the sheath at any time during the procedure.An icast 6x22 device was used to complete the procedure.There was no harm to the patient.There was no reported patient injury.
|
|
Manufacturer Narrative
|
It was reported that two devices could not be advanced through 6f introducer sheaths.There were no anomalies noted during the prep of the devices.There was no damage noted to the box, tray or stent cover and there was no difficulty removing the stent cover on either of the devices.The intended treatment site was the renals.This was a aaa procedure.The delivery system was flushed.Access was made via the iliac.A benson guidewire and 6f ansel cook introducer sheaths were used.The sheaths used had the same sheath product code.The vessel was pre-dilated using a 6x4 armada device.Two vascular covered stent delivery systems allegedly did not go through two 6fr introducer sheaths during an attempt to treat the renals via the iliacs.In both cases the device advanced about 2 inches into the sheath and then failed to advance further.The devices were removed back through the sheath.There was difficulty retracting the devices from the patient as it was reported that it was a little tight.Multiple attempts to insert the devices were not made.Air evacuation was performed prior to inserting both devices.Guidewire access was maintained throughout the procedure.An icast 6x22 device was used to complete the procedure.There was no harm to the patient.There was no reported patient injury.The lot history records have been reviewed with special attention to the manufacturing and inspection of this product.The device was found to have met specifications prior to shipment.No manufacturing anomalies were identified that may have caused or contributed to the reported event.This is the first reported complaint for this lot number and issue to date.The device was not returned for evaluation.Two identical boxes were returned for analysis.These boxes were returned without devices.The first box contained a 6f cook sheath and sheath's packaging.The second box contained two sheaths.These two sheaths returned were unidentified.There were no clear markings noted on either sheath.No evaluation could be performed as the sample was not returned.The result of the investigation is inconclusive.Based upon the available information and sample evaluation a definitive root cause has not been determined.It is unknown if patient factors, handling or procedural techniques may have contributed to the reported event.Based on analysis performed no additional action is required at this time.The ifu states: a device description: 1.Implant.The lifestream¿ balloon expandable vascular covered stent is comprised of an electropolished balloon-expandable stent made from 316l stainless steel, encapsulated between two layers of eptfe.B indication for use: the lifestream¿ balloon expandable vascular covered stent is indicated for the treatment of atherosclerotic lesions in common and external iliac arteries.Directions for use: site access and preparation; 1.Using standard techniques access the artery and place an introducer sheath or guiding catheter of appropriate inner diameter and a 0.035" (0.89 mm) guidewire across the target lesion.2.Perform diagnostic angiography to confirm site of implantation and measure the reference vessel diameter and lesion length.Covered stent size selection: 3.Select a covered stent diameter that is approximately 5%-20% larger than the largest reference vessel diameter at the proximal or distal target site.Refer to the sizing table on the packaging label for appropriate selection of the covered stent diameter and length.Endovascular system preparation: 4.Carefully remove the selected device from the package.5.Inspect the covered stent for adherence to the balloon and centered placement in relation to the balloon marker bands.If the covered stent is not centered and/or does not firmly adhere to the balloon, do not use.6.Flush the delivery system guidewire lumen with sterile saline mixture until saline drops from the distal end of the endovascular system.Air evacuation: 7.A 20 cc or smaller luer-lock syringe with a minimum of 5 cc¿s sterile saline mixture is recommended for use for aspirating this device.8.With the distal balloon tip pointing down and positioned below the level of the syringe, pull negative pressure until all air is expelled.9.Induce a negative pressure to remove any air from the balloon and inflation lumen.Repeat until all air is expelled.10.Carefully release to neutral.Allow the inflation lumen to fill with the diluted contrast medium and maintain a neutral pressure.Important: do not apply positive pressure to the balloon.11.Attach the prefilled inflation device to the inflation lumen of the catheter hub, ensuring no air bubbles remain at the catheter connection.12.Verify that the covered stent is still centered between the two radiopaque markers on the balloon catheter.Introduction of the endovascular system and placement of the covered stent 13.Advance the endovascular system over the guidewire into the introducer sheath.14.Further advance the endovascular system to the target treatment site within the introducer sheath and position the covered stent across the lesion.Verify that the covered stent is still centered within the balloon marker bands.Slowly retract the introducer sheath / guiding catheter while maintaining the position of the covered stent.Ensure the introducer sheath is retracted far enough to not compromise the balloon expansion and covered stent release.(b)(4).
|
|
Event Description
|
It was reported that two devices could not be advanced through 6f introducer sheaths.There were no anomalies noted during the prep of the devices.There was no damage noted to the box, tray or stent cover and there was no difficulty removing the stent cover on either of the devices.The intended treatment site was the renals.This was a aaa procedure.The delivery system was flushed.Access was made via the iliac.A benson guidewire and 6f ansel cook introducer sheaths were used.The sheaths used had the same sheath product code.The vessel was pre-dilated using a 6x4 armada device.Two vascular covered stent delivery systems allegedly did not go through two 6fr introducer sheaths during an attempt to treat the renals via the iliacs.In both cases the device advanced about 2 inches into the sheath and then failed to advance further.The devices were removed back through the sheath.There was difficulty retracting the devices from the patient as it was reported that it was a little tight.Multiple attempts to insert the devices were not made.Air evacuation was performed prior to inserting both devices.Guidewire access was maintained throughout the procedure.An icast 6x22 device was used to complete the procedure.There was no harm to the patient.There was no reported patient injury.
|
|