MAKO SURGICAL CORP. TIBIAL INLAY IMPACTOR HEAD; PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM
|
Back to Search Results |
|
Catalog Number 160177 |
Device Problems
Break (1069); Fracture (1260)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
Event Date 04/28/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
A supplemental report will be submitted upon completion of the investigation.
|
|
Event Description
|
It is reported that during a surgery, the impactor broke.During the first contact, the plastic part from impactor broke off.
|
|
Manufacturer Narrative
|
An event regarding crack/fracture involving a mako impactor was reported.The event was confirmed.Method & results: device evaluation and results: a visual inspection is performed and found that fracture consistent with an overload condition as indicated by hackles observed on fracture surface.Medical records received and evaluation: no medical records or x-rays were made available for evaluation.Device history review: device history review indicated the devices accepted into final stock from the reported lot were free from discrepancies complaint history review: there has been no other event for the lot referenced.Conclusions: the event is confirmed.A visual inspection of returned device is performed and found that fracture consistent with an overload condition as indicated by hackles observed on fracture surface.If the additional information are received, this investigation will be reopened and re-evaluated.
|
|
Event Description
|
It is reported that during a surgery, the impactor broke.During the first contact, the plastic part from impactor broke off.
|
|