MAUDE Adverse Event Report: MAKO SURGICAL CORP. TIBIAL INLAY IMPACTOR HEAD; PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAKO SURGICAL CORP. TIBIAL INLAY IMPACTOR HEAD; PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM Back to Search Results
Catalog Number 160177
Device Problems Break (1069); Fracture (1260)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 04/28/2017
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of the investigation.
 
Event Description
It is reported that during a surgery, the impactor broke.During the first contact, the plastic part from impactor broke off.
 
Manufacturer Narrative
An event regarding crack/fracture involving a mako impactor was reported.The event was confirmed.Method & results: device evaluation and results: a visual inspection is performed and found that fracture consistent with an overload condition as indicated by hackles observed on fracture surface.Medical records received and evaluation: no medical records or x-rays were made available for evaluation.Device history review: device history review indicated the devices accepted into final stock from the reported lot were free from discrepancies complaint history review: there has been no other event for the lot referenced.Conclusions: the event is confirmed.A visual inspection of returned device is performed and found that fracture consistent with an overload condition as indicated by hackles observed on fracture surface.If the additional information are received, this investigation will be reopened and re-evaluated.
 
Event Description
It is reported that during a surgery, the impactor broke.During the first contact, the plastic part from impactor broke off.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TIBIAL INLAY IMPACTOR HEAD
Type of Device
PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM
Manufacturer (Section D)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer (Section G)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer Contact
cindy chuhinko
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key6598441
MDR Text Key76275970
Report Number3005985723-2017-00233
Device Sequence Number1
Product Code NPJ
Combination Product (y/n)N
Reporter Country CodeLU
PMA/PMN Number
K090763
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 07/21/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number160177
Device Lot Number19040215
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/09/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/02/2017
Initial Date FDA Received05/30/2017
Supplement Dates Manufacturer Received06/22/2017
Supplement Dates FDA Received07/21/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/10/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-