• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES OBERDORF ATTACHM F/COMPACT AIR DRIVE+POWER DRIVE; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES OBERDORF ATTACHM F/COMPACT AIR DRIVE+POWER DRIVE; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT Back to Search Results
Catalog Number 511.785
Device Problem Detachment Of Device Component (1104)
Patient Problem No Patient Involvement (2645)
Event Date 05/04/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).(b)(6).This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.During repair, it was observed that the device was dirty, oily and corroded.Therefore, the reported condition was duplicated and confirmed.The assignable root cause was determined to be due to improper maintenance, which is user error/misuse/abuse.If additional information should become available, a supplemental medwatch report will be sent accordingly.
 
Event Description
It was reported from (b)(6) that the attachment device screw at the nose fell.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTACHM F/COMPACT AIR DRIVE+POWER DRIVE
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
Manufacturer (Section D)
SYNTHES OBERDORF
eimattstrasse 3
oberdorf CH443 6
SZ  CH4436
Manufacturer (Section G)
SYNTHES OBERDORF
eimattstrasse 3
oberdorf CH443 6
SZ   CH4436
Manufacturer Contact
michael cote
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key6602693
MDR Text Key76407409
Report Number8030965-2017-12667
Device Sequence Number1
Product Code HWE
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K971544
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 05/10/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/31/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number511.785
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/16/2017
Date Manufacturer Received05/10/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/22/2002
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-